LAI-2000 Specifications

LAI-2070 Control Unit:

Sensor Inputs: 2 DB15 connectors for LAI-2050 sensor heads (multiplexed, 5 channel). 2 BNC connectors for other LI-COR radiation sensors.

Memory: 64K bytes RAM for file storage. 16K bytes RAM for working memory. File size is user determined.

File Storage Capacity: Each file requires 250 bytes, plus 30 bytes per record. Memory capacity is 64K bytes but files cannot be larger than 16K or overlap a 16K boundary. So, for example, a file with 10 records (such as 2 above and 8 below canopy readings) takes up 250 + (10 x 30) = 550 bytes. 16K of memory will hold 29 of these files, so the entire 64K memory would hold 116 files.

Keyboard: 24 key tactile response keypad with audio feedback.

Display: 2 line by 16 character LCD.

Communications: RS-232 wired as Data Communications Equipment (DCE). UART supported, baud rates of 300, 1200, 2400, 4800; 7 or 8 data bits; parity = even, odd, none; xon-xoff either on or off; hardware output handshaking with DTR pin 20 required. FV2000 Windows software is included to download data from the LAI-2000 to a PC.

Clock: Year, Month, Day, Hour, Minute. Accuracy: ± 3 minutes per month.

Power Requirements: 6 alkaline "D" cell batteries.

Battery Life: 260 hours of continuous operation.

Low Battery Warning: Occurs when 15% of battery life remains.

Memory Backup Power: "D" cells power RAM for 1 year after automatic low voltage shut-off. RAM is also protected for short intervals (5 minutes) to facilitate battery changes.

 

LAI-2050 Optical Sensor:

Nominal Angular Coverage:
Ring 1: 0.0 - 12.3°
Ring 2: 16.7 - 28.6°
Ring 3: 32.4 - 43.4°
Ring 4: 47.3 - 58.1°
Ring 5: 62.3 - 74.1°

Optics: 1.00° maximum decentering error as measured from center of mass of ring 4. 0.50° maximum magnification error as measured from center of mass of ring 4.

Wavelength Range: 320-490 nm.

Radiation Rejection:
490-650 nm: to <1%.
>650 nm: to <0.1%.

Lens Coating: MgF2 for improved transmission at high oblique angles (front lens and others).

View Caps: Provides azimuthal masking into quadrants of 0°, 45°, 90°, 180°, and 270°.

Environmental Conditions: Operating: -15 to 50 °C, 0-100% RH (non-condensing). Storage: -30 to 60 °C, 0-100% RH.

Size:
Control Unit:
21 L x 11.4 W x 6.9 D cm (8.3" x 4.5" x 2.7").
LAI-2050 Optical Sensor: 61 L x 2.9 W x 2.9 D cm (24" x 1.125" x 1.125"). Handle is articulated 7" from lens and bends 90° down or 25° up.

Weight:
Control Unit:
1.8 kg (3.9 lb.).
LAI-2050 Optical Sensor: 0.85 kg (1.9 lb.).

Specifications subject to change without notice